Evaluation of Small Ultrasonic Probe using Lead Zirconate Titanate poly-crystal Film Deposited by Hydrothermal Method
The present study evaluates the characteristics of a small ultrasound probe using PZT polycrystalline film deposited by a hydrothermal synthesis method. With the hydrothermal synthesis method, PZT film is deposited only on the titanium by reaction in the solution. Utilizing this feature, a PTZ polycrystalline film was deposited on the end face by a hydrothermal synthesis of a material with only its titanium wire end face exposed. This transducer was built into a small ultrasound probe and the characteristics were evaluated. As a result, we ucceeded in the fabrication of a wideband ultrasound probe with a distance resolution of 32 μm, a frequency bandwidth of 50 MHz in the range of 25 to 74 MHz, and a fractional bandwidth of 99 %.
hydrothermal method PZT ultrasound probe evaluation
Tomohito HASEGAWA Shinichi TAKEUCHI
1Minoru Kuribayashi KUROSAWA Tokyo Institute of Technology Yokohama,226-8503,Japan Toin University of Yokohama Yokohama,225-8502,Japan
国际会议
西安
英文
332-336
2009-08-23(万方平台首次上网日期,不代表论文的发表时间)