Automatic Testing System for Parameters of High-current Thermoprotector
The testing system for parameters of high-current Thermoprotector uses PLC as its core .equipped with touch display screen and micro-printer.It also employs constant-current electronic load to supply constant high-current (100A). And utilizes hall element to convert current signal to voltage signal fed backed to PLC,which record the trip time or recovery time of thermoprotector .It can test whether the thermoprotectors insulation resistance is qualified,while the chattering can be detected. The system can test several parameters accurately and simply,and provide the function of data statistics,print and product sorting which enhance the work efficiency.
high-current thermoprotector constant-current electronic load trip time.
Yanhong Wei Chang Xu
College of Mechanical & Electrical Engineering,China Jiliang University,China
国际会议
2009 9th International Conference on Electronic Measurement & Instruments(第九届电子测量与仪器国际会议 ICEMI2009)
北京
英文
520-523
2009-08-16(万方平台首次上网日期,不代表论文的发表时间)