会议专题

opology Optimization In The Passive Shim for The Permanent Magnet NMR

An improved method for allowing a NMR magnet to meet its field homogeneity requirements has been developed.Using the adjoint variable method, topology design sensitivity is derived analytically. Coupled with continuum topology sensitivity analysis, steep descent method is employed to find an optimized distribution of the shimming block of iron for the permanent magnets NMR. Different free induction decay signal data shows the effectiveness of the proposed optimal design method.

NMR Permanent magnet Sensitivity analysis Topology optimization

Liu Yang Minghao Yang Lanlan Dong

College of Engineering,China Agricultural University,BeiJing,P. R. China 100083 College of Information and Electrical Engineering,China Agricultural University,BeiJing,P. R. China

国际会议

第三届亚洲精细农业会议暨第五届智能化农业信息技术国际会议

北京

英文

1-4

2009-10-14(万方平台首次上网日期,不代表论文的发表时间)