The valuation of uncertainty on cylinder shell resonating density meter
In order to improve the measurement accuracy of cylinder shell resonating density meter, the paper presents many means of methods to lower the measurement error from the liquid temperature, pressure, viscidity and gas bubble in the liquid, and designs a novel cylinder shell resonating density meter. Taking typical calibration experiments as example, the paper gives the detailed steps of calculating uncertainty component of the cylinder shell resonating density meter, and values the combined uncertainty. The experiments and analysis show the combined uncertainty is 0.96kg/m3 when the measurement result is 785.16 kg/m3, and the relative uncertainty is 0.12%.
Resonating sensor cylinder shell resonator density meter uncertainty valuation uncertainty component
Sui X Wa Li Y G X Fan Y M
Tianjin Key Laboratory of Advanced Mechantronics Equipment Technology, School of Machineryand Electr State Key Lab of Precision Measuring Technology & Instrument, Tianjin University,Tianjin,china,30007
国际会议
第五届仪器科学与技术国际学术会议(ISIST 2008)Fifth International Symposium on Instrmentation Science and Technology
沈阳
英文
1-7
2008-09-15(万方平台首次上网日期,不代表论文的发表时间)