会议专题

A Monte Carlo interstitial brachytherapy study for AAPM TG-43 dose calculation formalism in heterogeneous media

The precision of AAPM TG-43 dose calculation formalism in heterogeneous media has been studied based on the survey and evaluation of the status of interstitial brachytherapy. After a brief introduction of radioactive seeds (125I, 192Ir and 103Pd) and AAPM TG-43 formalism, the radial dose rates calculated by Monte Carlo code MCNP-4C and TG-43 formalism were compared with homogeneous and heterogeneous phantoms. The reason of errors between the results of the above two methods was analyzed, and modified dose rates constant were proposed and testified. With the updated dose rates constant, AAPM TG-43 formalism can predict the dose rate distribution accurately and rapidly in clinical.

Brachytherapy Radioactive seed AAPM TG- 43 Inhomogeneity Dose calculation

Gang Song Yican Wu

Institute of Plasma Physics, Chinese academy of Sciences Hefei, China

国际会议

The 2nd International Conference on Bioinformatics and Biomedical Engineering(iCBBE 2008)(第二届生物信息与生物医学工程国际会议)

上海

英文

1005-1008

2008-05-16(万方平台首次上网日期,不代表论文的发表时间)