Comparison of Drive Patterns for Single Current Source EIT in Computational Phantom

Electrical impedance tomography (EIT) estimated the electrical conductivity distribution within a body. Bipolar drive pattern appeared in most EIT studies that the current was injected with a single current source and sink. In the present study, we built a computational phantom and compared the quality of dynamic EIT image in terms of data quality, resolution and anti-noise performance associated with three different bipolar drive patterns: polar drive pattern, pseudo-polar drive pattern and adjacent drive pattern. The results demonstrated that the pseudo-polar drive pattern is optimal.

electrical impedance tomography drive pattern filter back projection

Canhua Xu Xiuzhen Dong Xuetao Shi Feng Fu Wanjun Shuai Ruigang Liu Fusheng You

Department of Biomedical Engineering Fourth Military Medical University Xian, China