Calibration for a one-dimensional synthetic aperture microwave radiometer is critical. A numerical image retrieving method using an intermediate transformation matrix called the gain matrix or simply G-matrix is considered to be a good method. This matrix takes into account all the couplings and un-uniformed amplitude and phase between the channels. However,measuring the G-matrix is not an easy job. One has to set up a point source in the far field and rotate the system in the fan beam direction. The difficulty is mainly due to the incomplete test environment. In this paper, a measurement set-up called image simulator is present. It is a circuit network composed of a number of high accurate microwave I-Q vector modulators and a noise generator. It simulates the incoming waves to the thinned antenna array corresponding to any image scene to be measured. The design and test of the image simulator are presented followed with test results of G-matrix. The image simulator can also simulate a point source in the two dimensional aperture case which discussed in the paper too.