New Design of Eutectic Fized-Point Cell for Simultaneous Contact and Non-contact Measurements

A new design of fixed point cell for simultaneous contact and non-contact measurements has been developed in MIRS/FE-LMK. The cell can be used in vertical or horizontal position. However it has been designed for use in horizontal furnaces where the temperature of the phase change transition can be measured from each side of the furnace tube, either with thermocouples or non-contact thermometers. The measurements were performed simultaneously with thermocouple and non-contact thermometer and also with two thermocouples simultaneously. Over 20 realizations of the Ni-C fixed point were performed with the developed cell. The melting temperatures of Ni-C measured with thermoeouple type R and non-contact thermometer agree within their uncertainties.

Metrology Metal-Carbon eutectic Fized point cell Thermocouple Radiation thermometer

Jovan Bojkovski Miha Hiti Janko Drnovsek

University of Ljubljana, Faculty of Electrical Engineering (MIRS/FE-LMK), Ljubljana, Slovenia

国际会议

2008年国际温度与热物性测量会议(TEMP BEIJING 2008)

北京

英文

38-41

2008-10-20(万方平台首次上网日期,不代表论文的发表时间)