会议专题

Contribution for a Very Accurate Determination of The Water Vapor Pressure Curve

In this work we start a study where the purpose is to investigate the relation between temperature and pressure in the case of pure water with the best uncertainty level in the temperature range between 173 K and 373 K. We present here our first step in this study, i.e. the first measurements, obtained thanks to a gas control heat pipe (GCHP), in the temperature range between 350 K and 385 K. These measurements are compared to data from literature and a preliminary uncertainty budget is proposed.

Metrology Vapor pressure curve Gas control heat pipe Phase equilibrium Pressure and temperature measurement

E.GEORGIN M.ELGOURDOU C.MARTIN

Centre des Industries Aérauliques et Thermiques LNE-CETIAT, 69100 Villeurbanne, France Institut National de Métrologie LNE -INM / CNAM, 93200 Saint Denis, France

国际会议

2008年国际温度与热物性测量会议(TEMP BEIJING 2008)

北京

英文

124-128

2008-10-20(万方平台首次上网日期,不代表论文的发表时间)