会议专题

Statistical Methods in Blackbody Sources Compare Calibration

Characteristic of Blackbody Sources Compare Calibration is investigated. The 3σ eliminate rule and arithmetic mean filter are exploited, which eliminate noise in Blackbody Sources Compare Calibration. The effect of statistical methods depresses the noise equivalent temperature difference (NETD) of radiation thermometer. Statistical methods can get the corresponding measurement level when choose different statistics quantity, the NETD could improved from 0.1 ℃ to better than 0.01 ℃. Statistical methods can be used not only in the Blackbody Sources Compare Calibration but also in the other compare calibrations as Blackbody Sources Compare Calibration.

Metrology Statistics method Noise Blackbody source Compare Calibration

WANG Jing-hui YUAN Zun-dong

National Institute of Metrology, Beijing 100013, China

国际会议

2008年国际温度与热物性测量会议(TEMP BEIJING 2008)

北京

英文

203-206

2008-10-20(万方平台首次上网日期,不代表论文的发表时间)