Improvement on Thermal Diffusivity Measurement Repeatability of Laser Flash Method
Electromagnetic interference (EMI) disturbs significantly thermal diffusivity measurements by laser flash method. This paper presented a progress in suppressing EMI by developing a new amplifier, programming a new software of digital filters, promoting the stability of furnace and applying a high performance data acquisition card. A armco iron sample whose thickness is 2.88mm was measured once again, the repeatability was observed to be improved from the previous 1%~3% into the present less than 8‰.
Metrology Laser flash method EMI Repeatability Labview
LIU Jian-qing SUN Jian-ping ZHANG Jin-tao CHENG Yue-fei
HeBei University, Baoding, Hebei 071002, China National Institute of Metrology, Beijing 100013, China Beijing University of Science and Technology, Beijing 100083, China
国际会议
2008年国际温度与热物性测量会议(TEMP BEIJING 2008)
北京
英文
280-282
2008-10-20(万方平台首次上网日期,不代表论文的发表时间)