会议专题

Neuro-Stimulus Chip with Photodiodes Array for Sub-retinal Implants

A prototype neuro-stimulus chip for sub-retinal implants in blind patients affected by Age-related Macular Degeneration (AMD) or Retinitis Pigmentosa (RP) is presented in this paper. This retinal prosthetic chip was designed to replace the degenerated photoreceptor cells,and in order to stimulate directly the remaining healthy layers of retinal neurons. The current stimulus circuits are monolithic integrated with photodiodes (PD) array,which can convert the illumination on the eyes into bi-phasic electrical pulses. In addition,a novel charge cancellation circuit is used to discharge the electrodes for medical safty. The prototype chip is designed and fabricated in HJTC 0.18 μrn N-well CMOS 1P6M Mixsignal process,with a± 2.5 V dual voltage power supply.

Xu Zhang Weihua Pei Beiju Huang Hongda Chen

State Key Laboratory on Integrated Optoelectronics,Institute of Semiconductor,Chinese Academy of Sciences,Beijing,100083

国际会议

9th International Conference on Solid-State and Integrated-Circuit Technology(第9届固态和集成电路国际会议)

北京

英文

1757-1760