会议专题

A CsI(Tl) detector array used in the ezperiment of the proton-rich nucleus 17Ne

To investigate the configurations of the valence protons in Borromean nucleus 17Ne, a CsI(Tl) detector array, which consists of 9 CsI crystals (26×26×20mm3) coupled with photodiodes, has been successfully used in the 17Ne experiment to measure the energy of protons. In order to find the optimal working conditions and get the best energy resolutions, several technologies (including various wrapping materials, wrapping and coupling methods) have been used. The testing results showed that the best energy resolution of the CsI(Tl) is about 3.3% using the 241 Am α-source. The primary testing results with the proton beam were also provided.

CsI(Tl), energy resolution, proton

MA Li-Ying HUA Hui LU Fei CHEN Dong JIANG Xi-Yao YE Yan-Lin JIANG Dong-Xing Qureshi Faisal-Jamil

School of Physics and State Key Laboratory of Nuclear Physics and Technology, Peking University, Beijing 100871, China

国际会议

第十二届全国核结构研讨会暨第八次全国核结构专题讨论会(12th National Conference on Nuclear Structure and the 8th Symposium on Nuclear Structure)

重庆

英文

176-178

2008-05-10(万方平台首次上网日期,不代表论文的发表时间)