Emissivity measurement study on wide aperture microwave radiator
The comparison research of emissivity measurement methods is presentedIn This paper. Different from the computation and measurement of general plane absorbing material,the emissivity measurement of pyramid microwave blackbody need consider near field effect,scattering effect and influence coming from measurement system reflection. Hence the measurement modeling of the former can not work well at measuring the emissivity of the latter. To improve emissivity measurement methods,the BIRMM (namely authors address)and the VNIIFTRI (All Russia scientific research institute for physical technical and radio technical measurement) finished together an international comparison measurement on same sample. The measurement methods used by BIRMM and VNIIFTRI and data are presentedInfollowing sections.
C.Y.CHENG F.Li Y.J.Yang Y.M.CHEN
Beijing Radio Institute of Metrology and Measurement,Beijing,100854,China National Key Laboratory of Beijing Radio Institute of Metrology and Measurement,Beijing,100854,China
国际会议
2008 International Conference on Microwave and Millimeter Wave Technology(2008国际微波毫米波技术会议)
南京
英文
914-917
2008-04-21(万方平台首次上网日期,不代表论文的发表时间)