会议专题

Intraoperative Local Demarcation System of Brain Tumor Based on 5-Aminolaevulinic Acid Induced Porphyrin Fluorescence

A lightweight intraoperative spectrum measurement system for 5-ALA induced PpIX fluorescence was developed.The developed system includes a semiconductor LASER for excitation light (output light wavelength:405nm),a CCD spectrometer,optical fiber coupled on lens system and PC.The first prototype device has a bundled fiber which contains 6 excitation fibers around one measurement fiber.The second one uses a dichroic mirror which reflects blue light (λ≤550nm) and transmits other wavelength.The third one contains not only a dichroic mirror but also a long path filter (cut on:λ=570nm) and has improved mechanical strength.The weight of the first and the prototype probe is 45g,and the external diameter is 11mm..The weight of this probe is 25g and the external diameter is 18mm.The probes can be positioned manually and measure 5-ALA induced PpIX fluorescence.

5-ALA Protoporphyrin IX Brain tumor Fluorescence measurement Handheld device

T.Ando I.Sakuma M.Noguch K.Shimaya K.Wang E.Kobayashi H.Liao T.Maruyama Y.Mulagaki H.Iseki

Graduate School of Engineering,The University of Tokyo,Japan Graduate School of Frontier Sciences,The University of Tokyo,Japan Department of Neurosurgery,Neurological Institute,Tokyo Womens Medical University,Japan Department of Neurosurgery,Neurological Institute,Tokyo Womens Medical University,Japan;Institute o

国际会议

7th Asian-Pacific Conference on Medical and Biological Engineering(第七届亚太地区生物工程学术会议)

北京