Current Densities inside Numerical Human Body Resulting from Low-Frequency Contact Currents

Current densities inside a numerical model of a human body due to low-frequency contact current are demonstrated in order to investigate the current density distribution inside the body.An anatomically-real Japanese adult male of the human numerical model is used,and the cubic voxel size is 2mm.In this paper,improved calculation method of current density is described,and results obtained when a skin conductivity σ=varies from 10-4 to 10-2S/m are shown.It is concluded that the contact current obtained for σ=10-2S/m is 129μA that is about five times as large as that obtained for σ=10-4S/m.

low-frequency contact current current densities skin conductivities

H.Tarao N.Hayashi K.Isaka

Takamatsu National College of Technology,Takamatsu,Japan Kyushu University,Kasuga,Japan The University of Tokushima,Tokushima,Japan