Electrical Parameters Measurement of Quartz Crystal in the π Network

The principle and its circuit implementation of quartz crystal parameters measurement are presented. The principle is based on the Communiqué of IEC 444 that gives the measuring method of electrical parameters of quartz crystal in a π network. There are two ways of measurement which are respectively the maximum transmission way and zero phase way. The impact on the measurement accuracy is analyzed theoretically and the hardware compensation methods of decreasing stray reactance are presented. The result of experiment proves that the error of series resonant frequency is less than ±2ppm ranging from 1MHz to 120MHz and the series resonant resistance can meet the design requirement.

πnetwork quartz crystal DDS

Liu Guili Wang Yanlin Li Dong

Department of Electronic Information Engineering, Beijing Institute of Machinery, Beijing, 100085 China

国际会议

第七届国际测试技术研讨会

北京

英文

2007-08-05(万方平台首次上网日期,不代表论文的发表时间)