会议专题

Parameter Optimization and Experiment of Chirp Excitation Ultrasound Imaging

Objective Parameters that affect image quality in Frequency-Modulated (FM) signal (viz. chirp) transmitting ultrasound imaging are investigated to increase Signal-to-Noise Ratio (SNR) of the present ultrasound image. Method The simulation and acoustical experiment are completed to investigate the image quality with various parameters of FM ultrasound signal transmitting. The energy of the transmitting ultrasound is increased, and the SNR is increased too. In order to ensure that the axial resolution of the image is not decreased, the convolution method is used to decode the received ultrasound signal. Result The transmitting parameters of FM signal transmitting ultrasound imaging were optimized. The phantom B-mode ultrasound image was gained using the method. The quality of the image is better than that of the traditional B-mode image. Conclusion The SNR of ultrasound image could be increased by FM signal transmitting method. The SNR of the image of deeper tissue could be increased about 5~8db, where the ultrasound signal is attenuated severely.

chip transmitting ultrasound imaging Signal-to-Noise Ratio1

GU Jinhong WANG Supin

Department of Biomedical Engineering, School of Life Science and Technology, Xian Jiaotong University, Xian, 710049

国际会议

第七届国际测试技术研讨会

北京

英文

2007-08-05(万方平台首次上网日期,不代表论文的发表时间)