会议专题

The Research of 3D-CT Image Reconstruction Algorithm ased on Projection Pre-procession

This paper presents a novel reconstruction algorithm for high transaxial-resolution cone-beam computed tomography (CT). In this paper, we propose projection pre-procession method that the projection data was filtered twice in high-frequency and low-frequency respectively with wavelet multi-scale decomposition, we call this algorithm frequency-split technique, which is designed to minimize cone-beam artifacts. Observing that cone beam artifacts contain mainly low-frequency components, the reconstructed image is a combination of two independent reconstructions with differently modified reconstruction filters. A high-frequency image is reconstructed with the ray projection data using a high-pass filter leading to a suppression of low-frequency artifacts, while the low-frequency components of the image are reconstructed by using low-pass filter, then add the two reconstructed images to make the high quality reconstructed image. The result of algorithm simulation showed the image quality improved and the system noise reduced apparently.

3-D reconstruction projection pre-procession wavelet decomposition

WANG Mingquan WANG Zhen

Key Laboratory of Instrumentation Science & Dynamic Measurement (North University of China), inistry of Education epartment of Information Engineering, North university of China, Taiyuan, 030051

国际会议

第七届国际测试技术研讨会

北京

英文

2007-08-05(万方平台首次上网日期,不代表论文的发表时间)