Research on Calibration Method of Magnetic Sensor in Hybrid-Source Magneto Tellurics

To obtain sensor sensitivity to the real weakmagnetic field intensity in field so that it can be used in Hybrid-source Magneto Tellurics, a kind of indoor calibration method of magnetic sensor is introduced. In this paper, with calibration conditions being satisfied, a calibration system was designed and results error was analysised. Because of the power-line disturbing, magnetic shielding and its influence to the calibrating precision are introduced.

Sensor sensitivity Sensor calibration Magneticshielding Mutual inductance

Wang Yanzhang Cheng Defu Wan Yunxia Lin Jun

College of Instrument Science & Electrical Engineering Ji Lin University,Changchun 130061,China;Texa College of Instrument Science & Electrical Engineering Ji Lin University,Changchun 130061,China

国际会议

第八届国际电子测量与仪器学术会议(Proceedings of 2007 8th International Conference on Electronic Measurement & Instruments)

西安

英文

2007-08-16(万方平台首次上网日期,不代表论文的发表时间)