会议专题

Development of a Computerized Multifunctional Form and Position Measurement Instrument

A model machine of multifunctional form and position measurement instrument controlled by a personal computer has been successfully developed. The instrument is designed in rotary table type with a high precision air bearing and the radial rotation error of the rotary table is 0.08 μm. Since a high precision vertical sliding carriage supported by an air bearing is used for the instrument, the straightaway motion error of the carriage is 0.3 μm/200 mm and the parallelism error of the motion of the carriage relative to the rotation axis of the rotary table is 0.4 μm/200 mm. The mathematical models have been established for assessing planar and spatial straightness, flatness, roundness, cylindricity, and coaxality errors. By radial deviation measurement, the instrument can accurately measure form and position errors of such workpieces as shafts, round plates and sleeves of medium or small dimensions with the tolerance grades mostly used in industry.

P Liu W Y Tian

School of Mechanical Engineering & Automation, Northeastern University, 11, 3 Lane, Wenhua Road, Shenyang 110004, China

国际会议

第四届仪器科学与技术国际会议( 4th International Symposium on Instrumentation and Science and Tcchnology)

哈尔滨

英文

566-570

2006-08-08(万方平台首次上网日期,不代表论文的发表时间)