Detection of Doppler Embolic Signals with Hilbert-Huang Transform

In order to identify the Doppler Ultrasound embolic signals for stroke prediction, it is important to confirm the embolic appearing time and frequency of movement. The paper proposes a new identification method based Hilbert-Huang Transform. The embolic signal is decomposed into a number of Intrinsic Mode Functions which are mono-component. Then Hilbert transform is applied to them to calculate the instantaneous frequencies and Hilbert amplitudes. The embolic appearing time can be confirmed according to the maximum amplitude which should be larger than the preset threshold. The corresponding instantaneous frequency can also be decided with the time-frequency distribution. The method is applied to 24 Doppler embolic signals. The time error for more than 95% Doppler signals is less than 70ms. The result shows the powerful potential of HHT will have a promising future in embolic signals identification and stroke prediction in the clinical application.

Doppler Ultrasound embolic signal Hilbert-Huang Transform (HHT) Hilbert Spectrum

Dongmei Hao Hongwei Zhang

School of Life Science and Bioengineering,, Beijing University of Technology, Beijing 100022, P. R. Instruments Center, NanKai Hospital, Tianjin 300100, P. R. China

国际会议

2007 IEEE/ICME International Conference on Complex Medical Engineering-CME2007(CME2007 第二届国际复合医学工程学术大会)

北京

英文

413-416