会议专题

A Resonator for a Wrist-Sized, 0.2T Permanent Magnet MRI Machine

Low field MRI machines are important, since the machines offer easy access for users. However, the SNR attained with a low field MRI machine is not necessarily high. This paper attempts to present an effective resonator coil designed in particular for a human hand placed in a wrist-sized permanent magnet MRI machine. The optimal design of a solenoidal coil is investigated. The wire conductor features and the proximity to the other conductors are also investigated. A rectangular solenoidal coil was built. MR Images of the hand were obtained. The wrist contour, finger muscles, fatty tissues and bone marrow were distinctly displayed.

M. Murao T. Miyamoto M. Migita K. Ogawa H. Akiyama T. Usagawa N. Iriguchi

Kumamoto University Faculty of Engineering and Graduate School, Kumamoto, Japan Keio University Graduate School of Engineering, Yokohama, Japan

国际会议

2007 IEEE/ICME International Conference on Complex Medical Engineering-CME2007(CME2007 第二届国际复合医学工程学术大会)

北京

英文

864-867

2007-05-23(万方平台首次上网日期,不代表论文的发表时间)