Real-time High Resolution Optical Coherence Microscopy with a Phase Shifting Apodizer
In conventional optical coherence microscopy (OCM) systems, there is an inherent tradeoff between the depth of focus (DOF) and lateral resolution. A new method to overcome this limitation for real-time optical coherence microscopy is proposed. By simply incorporating a three-zone, pure phase, phase shifting apodizer in a conventional OCM setup, we improve DOF by a factor of 3. A nearly constant lateral resolution of 5.0 μm or better over a depth range of 500 μm is experimentally verified. Since pure phase masks can be easily fabricated and miniaturized, the merits of the proposed method are especially desired in real-time, high-resolution endoscopic OCM systems.
Linbo Liu Cheng Liu Wong Chee Howe C.J.R. Sheppard Nanguang Chen
1Division. of Bioengineering, National University of Singapore, Singapore, 117576;Graduate Programme 1Division. of Bioengineering, National University of Singapore, Singapore, 117576 Department of Electrical and Computing Engineering, National University of Singapore, Singapore, 117 1Division. of Bioengineering, National University of Singapore, Singapore, 117576;Graduate Programme
国际会议
北京
英文
1111-1114
2007-05-23(万方平台首次上网日期,不代表论文的发表时间)