Study on The Energy Loss Model of Ultrasonic Surgery Scalpel

Ultrasonic transducer is the kernel part of ultrasonic vibration system, changes electric energy into mechanical energy and determines the performance of system mostly. To reduce the temperature of ultrasonic surgery scalpel, it is necessary to analyze the energy loss of the ultrasonic transducer. First, The energy loss model of the thin equation section rod is deduced based on the theory of electrical transmission line and two ports network, while taking none but internal friction into account. Then, the energy loss model of the longitudinal direction ultrasonic transducer is established based on the theory of electrical transmission line and Mason model. The difference of maxim import conductance between the theory calculated value and experiment value is 0.065S in the experiment. The main reason is taking no structure loss of transducer into accounted. Compared with the traditional plural constant model, this method can easily calculate the energy loss of the ultrasonic transducer as long as the material mechanical quality factor is given.

Qun Cao Xiuzhen Li Chunyuan Li Qin Li

School of Life Science & Technology, Beijing Institute of Technology, Beijing, 100081, P. R. China Beijing StarPoint Communication Software Co., Ltd, Beijing, 100088, P. R. China

国际会议

2007 IEEE/ICME International Conference on Complex Medical Engineering-CME2007(CME2007 第二届国际复合医学工程学术大会)

北京

英文

1366-1369

2007-05-23(万方平台首次上网日期,不代表论文的发表时间)