Calculation and Simulation of Temperature in Real-time PCR Instrument
In this paper, we designed a model of the PCR reaction chamber by ourselves and calculated the heat flux of the reaction chamber, the lid and the sample. On the basis of the temperature calculation we simulated and analyzed the temperature fields dynamic distribution of the reaction system and we find that the simulative result and the calculating result match each other. This offers an academic foundation to develop a self-design real-time PCR instrument.
C. B. Liu J. H. Feng
Biomedical Engineering Centre, Beijing University of Technology, Beijing, 100022, China
国际会议
北京
英文
1597-1600
2007-05-23(万方平台首次上网日期,不代表论文的发表时间)