Determination of gibberellins by flow injection analysis based on chemiluminescence system
A novel method for determination of gibberellins (GAs) using flow-injection chemiluminescence analysis was described in this paper.The chemiluminescence intensity of the solution was enhanced proportionally while the concentration of gibberellins increased.Under the selected experimental conditions,the calibration curve of gibberellins was linear within the range from 5.0×10-3 to 1.0×10-1 mg.mL-1 with a linear equation of △I = 7098.5×(mg.mL-1) + 87.4,R2=0.9986.The detection limit (3σ)was 4.23×10-4 mg.mL-1.In the case of continuous determination of gibberellins,the relative standard deviation (RSD,n=11) was less than 2.31 %.The method has been successfully applied to the determination of industrial production with satisfactory results.