Determination of Metal Ions in Water Samples by Capillary Electrophoresis with Indirect UV Method

  A simple indirect UV method was developed for the separation of ten trace metal ions by capillary electrophoresis.The background electrolyte contains 15 mmol/L imidazole as the chromophore for indirect UV detection, and acetic acid functioned as a pH modifier for adjusting pH to 3.7 and weak complex reagent.All cations were separated completely within 8 min under hydrodynamic injection mode with a running voltage of +18 kV at 25 ℃.The limits of detection ranged from 0.02 mg/L to 0.21 mg/L and relative standard deviations (n =5) were in the range from 2.1% to 4.2%.The minimum quantitative levels located in the range between 0.10 mg/L and 0.47 mg/L and the maxima between 30 mg/L and 60 mg/L.The recovery for standards and actual water samples was also investigated respectively.The results demonstrate that the method is convenience and reproducible for practical analysis of water systems.

capillary electrophoresis metal ions indirect method water samples

Mei SHI Qingyu GAO Yongchao LU Jiamin FENG

School of Chemical Engineering and Technology, China University of Mining and Technology, Xuzhou, Jiangsu, 221116, P.R.China

国内会议

2010中国可再生能源科技发展大会

北京

英文

475-479