The Amines and Phenols Existing in RFCC Gasoline Affect the Accuracy of Potentiometric Titration

This paper described that there were two or more endpoints in the curve when potentiometeric titration”1” was used to determine the contents of mercaptan sulfur in residue fluidized catalytic cracking (RPCC) gasoline. Comparing with the one endpoint of a mixture solution of C2 - SH to C10 - SH determined under the same conditions, we found some components in RFCC gasoline affected the result of titration. A facile method washing the gasoline with 2% HC1 solution was brought out to eliminate the interfering components from the RFCC gasoline.After being treated with this method, all RFCC gasoline that contained 14 to 27 μg/mL mercaptan sulfur (SRSH) met the quality specification (SRSH shall not exceed 10μg/mL )”2”. The analysis of gas chromatography mass spectrometry (GC - MS) and gas chromatography infrared spectrometer (GC- IR) verified that the interfering components were aniline, phenol and their alkyl - substituted derivatives.

RFCC gasoline mercaptan potentiometric titration GC-MS GC-IR

Xingguo Cheng Liming Huang Peixun Li Zhide Hu Lei Cao

Department of Chemistry, Lanzhou University, Lanzhou 730000, P. R. China Lanzhou lube oil R&D insti Department of Material science, Lanzhou University, Lanzhou 730000, P.R. China Department of Chemistry, Lanzhou University, Lanzhou 730000, P. R. China Lanzhou lube oil R&D institute of Petrachina,Lanzhou 730060, P.R. China