SPECTROPHOTOMETRIC METHOD FOR THE DETERMINATION OF IRON SPECIES IN THE PRESENCE OF FERRIC IRON
The high concentration of ferric(Fe3+)in sample has interference on the determination of ferrous(Fe2+)concentration by spectrophotometric method which 1,10-phenanthroline is used to as chromogenic agent.Appropriate concentration of fluorine ion(F-)can complex Fe3+to eliminate interferences which the F-concentration is 13.3 time greater than that of Fe3+.The temperature and light do not affect the masking action of F-.Low temperature and dark conditions favour the stability of chromophoric complex.This method is suitable for determination iron species in the presence of high concentration of Fe3+in sample.
Iron species Spectrophotometricmethod Fluorine ion Masking action Storage condition
Feifei Fan Xiaoyun Guan Yitong Li Jian Zhu
Guizhou Institute of Soil and Fertilizer,Xiaohe District,Guiyang,550006,P.R.China School-Enterprise Office of Yibin Vocation and Technical College,Cuiping District,Yibin,644003,P.R.C College of Resource and Environmental Engineering,Guizhou University,Huaxi District,Guiyang,550025
国际会议
2017 5th International Symposium on Computer and Electrical Engineering(2017年第五届计算机与电气工程国际学术研讨会)
长沙
英文
1-3
2017-10-29(万方平台首次上网日期,不代表论文的发表时间)