Performance Study of SMPS and FDMA for Nanometer Aerosol Particle Measurements

A new method of Differential Mobility Analyzer(DMA) operation using a Fast Scanning Differential Mobility Analyzer(FDMA)nanometer particle sizer has been implemented.By optimizing the flow and voltage ramps,improvements can be made in the FDMA’S measurement size range,counting statistics and resolution.In this study,an electrostatic classifier was used to generate mono-dispersed particles ofdifferent sizes to calibrate the delay time of the FDMA.The aerosol from incense combustion could be measured directly by FDMA although the particulate concentration was as high as 1,011 particles cm-3.The resolution of the FDMA was 3 nm and better than the SMPS (10nm).
Fast-scan differential nanometer analyzer organic aerosol inorganic aerosol SMPS nanometer analyzer
Chang Tang Chang Chih Ming Ma Gui Bing Hong Ham Sun Huang Da Ren hen
Department of Environmental Engineering,National I-Lan University,No.1 Sheen-Long Road,I-Lan City,26 Department of Cosmetic Application and Management,St.Mary’S Medicine Nursing and Management College, Graduate Institute of Environmental Engineering,National Taiwan University,71 Chou-Shan Rd.,Taipei 1 Department of Enemy,Environmental and Chemical Engineering,Washington University in St.Louis,St.Loui