Result Analysis on Quality Control Test for X-ray Computed Tomography Scanners from the Second Level 2nd Class hospitals in Shanghai
Objective The purpose of this study is to investigate and assess the present situation about the overall performance quality of CT scanners from Second Level 2nd Class hospitals.Methods High-contrast resolution, low contrast detectability, mean CT number of water, noise and uniformity were measured with a Catphan(R) 500 CT Phantom made by American Phantom Laboratory.A total of 62 CT scanners from 59 hospitals were tested.